Time to -30°C as a predictor of acute success during cryoablation in patients with atrial fibrillation.
Autores de ISABIAL
Participantes ajenos a ISABIAL
- Álvarez-Ortega CA
- Ruiz MA
- Solórzano-Guillén C
- Barrera A
- Toquero-Ramos J
- Martínez-Alday JD
- Grande C
- Segura JM
- García-Alberola A
- Moriña-Vázquez P
- Ferrero-de-Loma-Osorio Á
- Villuendas R
- Rocío C
- Arcocha MF
- Peinado R
Grupos de Investigación
Abstract
BACKGROUND: Freezing rate of second-generation cryoballoon (CB) is a biophysical parameter that could assist pulmonary vein isolation. The aim of this study is to assess freezing rate (time to reach -30°C ([TT-30C]) as an early predictor of acute pulmonary vein isolation using the CB. METHODS: Biophysical data from CB freeze applications within a multicenter, nation-wide CB ablation registry were gathered. Successful application (SA), was defined as achieving durable intraprocedural vein isolation with time to isolation in under 60 s (SA-TTI<60) as achieving durable vein isolation in under 60 s. Logistic regressions were performed and predictive models were built for the data set. RESULTS: 12,488 CB applications from 1,733 atrial fibrillation (AF) ablation procedures were included within 27 centers from a Spanish CB AF ablation registry. SA was achieved in 6,349 of 9,178 (69.2%) total freeze applications, and SA-TTI<60 was obtained in 2,673 of 4,784 (55.9%) freezes and electrogram monitoring was present. TT-30C was shorter in the SA group (33.4 ± 9.2 vs 39.3 ± 12.1 s; p < 0.001) and SA-TTI<60 group (31.8 ± 7.6 vs. 38.5 ± 11.5 s; p < 0.001). Also, a 10 s increase in TT-30C was associated with a 41% reduction in the odds for an SA (odds ratio [OR] 0.59; 95% confidence interval [CI] 0.56-0.63) and a 57% reduction in the odds for achieving SA-TTI<60 (OR 0.43; 95% CI 0.39-0.49), when corrected for electrogram visualization, vein position, and application order. CONCLUSIONS: Time to reach -30°C is an early predictor of the quality of a CB application and can be used to guide the ablation procedure even in the absence of electrogram monitoring.
Datos de la publicación
- ISSN/ISSNe:
- 1898-018X, 1898-018X
- Tipo:
- Article
- Páginas:
- -
- PubMed:
- 34708863
Cardiology Journal
Citas Recibidas en Web of Science: 4
Documentos
- No hay documentos
Filiaciones
Keywords
- atrial fibrillation; atrial fibrillation ablation; cryoballoon ablation; pulmonary vein isolation; second-generation cryoballoon
Cita
Álvarez Ortega CA, Ruiz MA, Solórzano Guillén C, Barrera A, Toquero Ramos J, Martínez Alday JD, Grande C et al. Time to -30°C as a predictor of acute success during cryoablation in patients with atrial fibrillation. Cardiol J. 31 agosto 2023;https://doi.org/10.5603/CJ.a2021.0135.
Portal de investigación