Failure of the new underexpansion criteria to predict outcomes after ACURATE neo2 valve implantation: results from an independent multicenter registry.

Fecha de publicación: Fecha Ahead of Print:

Autores de ISABIAL

Participantes ajenos a ISABIAL

  • Pan M
  • González-Manzanares R
  • Blas SG
  • Torralba EI
  • Molina AA
  • de la Torre-Hernández JM
  • Tébar D
  • Gallo I
  • Rivadulla C
  • Saura FT
  • Menchero AEG
  • Díaz JF
  • Fernández GV
  • Vila-García J
  • Moreno R
  • Sanchis J
  • Ojeda S

Grupos de Investigación

Abstract

INTRODUCTION AND OBJECTIVES: Suboptimal outcomes with the ACURATE neo2 valve have been attributed to valve frame underexpansion, reported in approximately 20% of cases according to 3 recently proposed angiographic criteria. We aimed to validate these criteria in an independent registry with a 2-year follow-up. METHODS: This registry included 696 patients with severe aortic stenosis treated with the ACURATE neo2 valve at 7 experienced centers. Underexpansion was defined as the presence of at least 1 of the 3 criteria; the boundary box method was also applied to assess postparallelism. The primary endpoint was a composite of all-cause mortality, stroke, or heart failure hospitalization at 2 years. RESULTS: The mean age was 81 ± 5.8 years. Based on the 3-criteria method, underexpansion was identified in 154 patients (22%). There were no significant differences in the primary endpoint between the 2 groups. The cumulative incidence of the composite outcome was 13.7% vs 11.0% at 1 year and 21.0% vs 17.4% at 2 years for the underexpanded and expanded groups, respectively (P = .535). Similarly, no differences were observed with the boundary box method. On multivariable analysis, underexpansion was not independently associated with the primary endpoint (HR, 1.15; 95%CI, 0.73-1.80, P = .537). The only variables significantly associated with major adverse events at 2-years were atrial fibrillation, diabetes mellitus, and reduced glomerular filtration rate. CONCLUSIONS: The presence of at least 1 angiographic underexpansion criterion after ACURATE neo2 implantation did not predict adverse outcomes at follow-up.

Copyright © 2025 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.

Datos de la publicación

ISSN/ISSNe:
1885-5857, 1885-5857

REVISTA ESPANOLA DE CARDIOLOGIA  EDICIONES DOYMA S A

Tipo:
Article
Páginas:
397-405
PubMed:
41038446
Enlace a otro recurso:
doi.org

Citas Recibidas en Web of Science: 1

Documentos

  • No hay documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • ACURATE neo2; Aortic stenosis; Estenosis aórtica; Infraexpansión; TAVI; Underexpansion

Cita

Compartir