A Novel Device for Temporary Cardiac Stimulation Using Permanent Pacemaker Generators and Active-Fixation Leads: The TEMPACE Multi-Center Prospective Study.

Fecha de publicación: Fecha Ahead of Print:

Autores de ISABIAL

Participantes ajenos a ISABIAL

  • Rivas-Gándara N
  • Jordán Marchite P
  • Adeliño R
  • Carmona MÁ
  • Bertomeu-Gonzalez V

Grupos de Investigación

Abstract

AIMS: To determine the safety and effectiveness of the KronoSafe® permanent pacemaker adapter for temporary cardiac pacing (TCP) with active-fixation leads (TCPAFL). METHODS: This was a multicenter, prospective, descriptive clinical investigation involving a medical device (ClinicalTrials.gov Identifier: NCT05351658). Between January 2023 and December 2024, all consecutive patients who underwent implantation of a temporary active-fixation pacemaker were included. The devices were secured using the KronoSafe® adapter throughout the period of use. Complications during TCPAFL were recorded, and R-wave sensing, lead impedance, and pacing threshold were determined every 48 hours. RESULTS: Thirty patients (10 with atrioventricular block, 9 TAVI, 5 alcohol septal ablations, 2 slow atrial fibrillations, 2 bradycardia-tachycardia syndrome, and 2 pacemaker infections) who required TCP were included. Two centers participated in the study. The mean duration of TCP was 7.8 days (maximum 22 days), with 79.2% of the time spent in the cardiology ward. One complication (3%) was recorded, due to accidental traction of the system associated with an episode of agitation. CONCLUSIONS: TCPAFL using the KronoSafe® adapter is safe and effective, and its use can be extended outside the Intensive Care Unit.

© The Author(s) 2025. Published by Oxford University Press on behalf of the European Society of Cardiology. All rights reserved. For commercial re-use, please contact reprints@oup.com for reprints and translation rights for reprints. All other permissions can be obtained through our RightsLink service via the Permissions link on the article page on our site—for further information please contact journals.permissions@oup.com.

Datos de la publicación

ISSN/ISSNe:
2048-8726, 2048-8734

European Heart Journal-Acute Cardiovascular Care  OXFORD UNIV PRESS

Tipo:
Article
Páginas:
732-738
PubMed:
40875828
Enlace a otro recurso:
doi.org

Citas Recibidas en Web of Science: 1

Documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • External Permanent Pacemaker; Temporary Pacemaker; Temporary Pacemaker with Active-Fixation Leads; Temporary-Permanent Pacemaker

Cita

Compartir