Efficacy and safety of novel left ventricular pacing leads: 1-year analysis of the NAVIGATOR trial

Fecha de publicación: Fecha Ahead of Print:

Autores de ISABIAL

Participantes ajenos a ISABIAL

  • De Sousa, J
  • Dompnier, A
  • Martins-Oliveira, M
  • Israel, CW
  • Teijeira, E
  • Rubin, JM
  • Sebag, F
  • Martino, M
  • Michel, Y
  • Marques, P

Grupos de Investigación

Abstract

Objectives Assess safety and performance of novel quadripolar preshaped left ventricular (LV) leads: NAVIGO 4LV 2D ('S shaped') and NAVIGO 4LV ARC ('U shaped').Methods Patients indicated for cardiac resynchronisation therapy were enrolled in a multicentre, prospective, controlled study (NAVIGATOR, NCT03279484). Patients were implanted with either a NAVIGO 4LV 2D or ARC lead, and assessed at 10 weeks, 6, 12 and 24 months post-implant. Co-primary safety and performance endpoints were assessed at 10 weeks. Safety endpoint was the patients' rate free from lead-related complications. Performance endpoint was the rate of patients with successful lead performance, defined as LV pacing threshold <= 2.5 V at 0.5 ms on at least one pacing vector, and the absence of phrenic nerve stimulation at the final programmed configuration. Lead-related complications and electrical parameters were monitored throughout study.Results A NAVIGO 4LV lead was successfully implanted in 211 out of 217 patients (97.2%). The safety endpoint was met, with 100% and 96.1% of patients free from complications for NAVIGO 4LV 2D and ARC, respectively. The performance endpoint was met with 98.1% and 98.9% of patients with a successful lead performance for NAVIGO 4LV 2D and ARC, respectively. Over 12 months, the global complication-free rate for both leads was 97.1% (95% CI: 93.71% to 98.70%), with a mean pacing capture threshold of 1.23 V +/- 0.73 V and a mean impedance of 951 ohm +/- 300.1 ohm.Conclusion A high implantation success rate and low complication rate was reported for the novel NAVIGO 4LV 2D and ARC leads, along with successful performance up to 12 months.

© Author(s) (or their employer(s)) 2024. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.

Datos de la publicación

ISSN/ISSNe:
2053-3624, 2053-3624

Open Heart  BMJ PUBLISHING GROUP

Tipo:
Article
Páginas:
-
PubMed:
38316493
Enlace a otro recurso:
openheart.bmj.com

Citas Recibidas en Web of Science: 1

Documentos

Métricas

Filiaciones mostrar / ocultar

Keywords

  • cardiac resynchronization therapy; LV lead; complication; safety